Falcon 2 (fast alcohol tester)   FS-LY-2       
    Product Description:
  • Front
  • Front
  • 1
  • 2
1, using high-performance processor, processing power.
2. It adopts a new generation of imported electrochemical sensors, which has the characteristics of fast running speed, high precision, good stability and long service life.
3, screen display: color OLED display, the screen background color automatically switches under different lighting conditions, to ensure that the screen is also clear under heavy light, the alcohol test process is dynamically displayed on the screen, convenient for duty.
4, waterproof function: all buttons and USB interface have a simple and suitable waterproof design;
5. The functions of forensics and rapid screening are combined into one;
6. Unique bell mouth air intake head design, with guiding and collecting gas function, ensuring more accurate screening and detection function;
7, traffic baton and flashlight lighting function, more convenient traffic police on duty;
8, built-in Bluetooth module, can be used with a variety of Bluetooth devices, dot matrix printers and thermal printers;
9, can store 80,000 test records, all records can be backed up to the background computer;
10. Adopt special design anti-back suction one-way blow pipe to ensure the safety and safety of the measurement process;
11. Vibration motor prompt function: When the test result is lower than the national standard, the vibration motor does not vibrate, the indicator light does not work; when the test result is higher than the national standard, the vibration motor emits a vibration prompt, the indicator light flashes, allowing the tester to be detected. People know in time whether there is alcohol in the test results.
12. Quick inspection and detection: The motor is used for sampling and pumping. The person being tested can automatically sample and analyze as long as it is close to the air inlet, without manual button sampling. The detection is fast and the result is 1 second.
13. The whole process of instructing the blowing process, the blowing time and flow monitoring function, intuitive

Products Demo Photo:

Technical Datas:

Measuring range 0mg/L~2.86mg/L
Indication error Alcohol gas concentration range C (mg/L)
C<0.400 ±0.020mg/L
0.400≤C<1.000±5%
C≥1.000±20%
Reproducibility Alcohol gas concentration range C (mg/L)
C<0.400 -0.007mg/L
0.400≤C<1.000-1.75%
C≥1.000-6%
Calibration cycle 6 months
Range of working temperature -5℃~45℃
Storage temperature range -30℃~70℃
Benchmark working environment conditions Temperature: 20 ℃ ± 5 ℃, humidity: <80% RH Atmospheric pressure: 86kPa~106kPa around there is no gas and electromagnetic field that affect the normal operation of the instrument
Display OLED digital display
Environmental pressure 600~1400hPa
Environment humidity 20~98%r.h.
Battery 1 rechargeable battery, 3.7V/3000mAh
Continuous measurement time 8 hours or more, flashlight
Host size ≤330 (mm)× outer diameter of about 45 (mm)
Instrument weight ≤270 g (with battery)
Test method Test distance 4-5cm test, test concentration (mg/100ml) 20, 40, 150
Allowable error(mg/100ml) Sensitivity
Sensitivity Test distance 8-10cm test, test 20mg/100ml concentration, instrument measurement value is greater than 8mg/100ml
Pumping mode The motor is continuously pumped, the pumping speed is not less than 0.8L/m